PEEK as an Alternative Biomaterial in Prosthodontics: Current Evidence and Future Perspectives
Department of Prosthodontics and Crown & Bridge, K.D. Dental College & Hospital, Mathura, UP, India
ABSTRACT
Polyetheretherketone (PEEK) has emerged as one of the most extensively investigated high-performance thermoplastic polymers in restorative and reconstructive dentistry over the past decade. Its favourable strength-to-weight ratio, chemical inertness, radiolucency, and an elastic modulus approximating that of cortical bone have positioned it as a metal-free and all-ceramic alternative across multiple prosthodontic domains, including fixed and removable partial denture frameworks, implant-supported prostheses, complete denture bases, and maxillofacial rehabilitation. This review synthesises the current understanding of PEEK's chemistry, mechanical behaviour, manufacturing pathways, and surface-modification strategies, and appraises the existing clinical and pre-clinical evidence for its prosthodontic applications. Persisting challenges — chiefly its hydrophobic, low-energy surface that limits adhesive bonding, the need for veneering to achieve acceptable esthetics, and the paucity of long-term randomised clinical trials — are discussed alongside emerging solutions such as high-performance PEEK composites (e.g., ceramic-reinforced and carbon-fibre-reinforced formulations), additive manufacturing, and novel surface-functionalisation protocols. While short- to medium-term data are encouraging, well-designed prospective clinical trials with extended follow-up are required before PEEK can be regarded as a routine, evidence-based substitute for conventional metallic and ceramic prosthodontic materials.
Keywords: Polyetheretherketone; PEEK; biomaterials; prosthodontics; dental implants; removable partial denture; surface treatment; digital dentistry
INTRODUCTION
The search for an ideal prosthodontic material, one that combines adequate mechanical strength, biocompatibility, favourable esthetics, and biomechanical compatibility with adjacent hard and soft tissues has driven continuous innovation in dental materials science. Conventional options such as cobalt-chromium alloys, titanium, and zirconia, while clinically proven, are associated with recognised limitations: metal hypersensitivity, corrosion and ion release, radiopacity that can obscure imaging, high elastic modulus leading to stress concentration and stress shielding at the bone-implant or tooth-restoration interface, and, in the case of zirconia, brittleness and technique-sensitive bonding protocols.1,2
Polyetheretherketone (PEEK) is a linear, semi-crystalline aromatic polymer belonging to the polyaryletherketone (PAEK) family, first introduced for industrial and aerospace use in the 1980s before being adopted in orthopaedic surgery most notably for spinal fusion cages owing to its radiolucency and bone-like stiffness.1 Its transition into dentistry over the past 10–15 years has been accelerated by the parallel maturation of CAD/CAM milling and, more recently, additive manufacturing, which allow precise, reproducible fabrication of PEEK-based restorations and frameworks.2,3 For the prosthodontic postgraduate, PEEK is of particular interest because it intersects nearly every domain of the specialty: fixed prosthodontics (frameworks, provisional restorations), removable prosthodontics (RPD frameworks and clasps, complete denture bases), implant prosthodontics (abutments, bar and hybrid frameworks), and maxillofacial rehabilitation.3,4
This review aims to consolidate the current evidence on the chemistry, mechanical and biological properties, manufacturing, and clinical performance of PEEK in prosthodontics, and to critically examine the gaps that must be addressed before it can be considered a mainstream, evidence-based alternative to established materials.
CHEMICAL STRUCTURE AND MATERIAL PROPERTIES
PEEK consists of repeating para-substituted phenyl rings linked by ketone and ether functional groups. This aromatic backbone confers high thermal stability (glass transition temperature ≈143°C, melting point ≈343°C), resistance to hydrolysis, and chemical inertness to acids, bases, and organic solvents encountered in the oral environment.1,2 Unlike metals, PEEK does not corrode or release ions, and unlike some ceramics, it does not fracture catastrophically, instead exhibiting a degree of plastic deformation before failure.
The property most frequently cited as clinically advantageous is its elastic modulus of approximately 3–4 GPa for unfilled PEEK, which lies closer to that of cortical bone and dentin than titanium (~110 GPa) or zirconia (~200 GPa).4,5 This similarity is proposed to reduce stress-shielding at the implant-bone interface and to provide a damping effect that favourably redistributes occlusal forces, though the flexural strength of unfilled PEEK (150–170 MPa) is correspondingly lower than that of metals and ceramics, limiting its use as a monolithic material in high-load areas.5,6
To address this trade-off, reinforced PEEK composites have been developed. Carbon-fibre-reinforced PEEK (CFR-PEEK) can achieve an elastic modulus of up to 18 GPa while retaining a favourable strength-to-stiffness ratio, though the exposed dark fibres compromise esthetics and are generally reserved for substructures.[4] Ceramic-filled PEEK formulations (e.g., commercially available as high-performance polymer systems reinforced with approximately 20% ceramic filler) raise the elastic modulus modestly to the 3–4 GPa range while improving surface hardness, wear resistance, and dimensional stability, and have been increasingly used for crowns, fixed partial denture frameworks, bars, and hybrid prostheses. A comparative summary of key mechanical properties relative to conventional prosthodontic materials is presented in Table 1.
Table 1. Comparative mechanical and physical properties of PEEK and conventional prosthodontic materials
|
Property |
PEEK (unfilled) |
CFR-PEEK |
Titanium (Ti-6Al-4V) |
Zirconia (Y-TZP) |
Cobalt-Chromium |
|
Elastic modulus (GPa) |
3–4 |
18 |
110 |
200–210 |
180–220 |
|
Density (g/cm³) |
1.30–1.32 |
1.4–1.6 |
4.43 |
6.05 |
8.5–9.2 |
|
Flexural strength (MPa) |
150–170 |
250–300 |
860–900 |
900–1200 |
600–700 |
|
Radiopacity |
Radiolucent |
Radiolucent |
Radiopaque |
Radiopaque |
Radiopaque |
|
Corrosion / ion release |
None (inert) |
None (inert) |
Possible |
None |
Possible (Ni/Co) |
|
Approx. relative cost |
Moderate |
Moderate–high |
High |
High |
Moderate |
Additional properties relevant to prosthodontic use include low specific gravity (approximately one-fifth that of titanium), which contributes to lighter, more comfortable prostheses; inherent radiolucency, which is advantageous for postoperative imaging but necessitates radiopaque markers in some applications; and a naturally tooth-like, ivory-white to grey colour that, while more esthetic than metal, generally still requires veneering with composite resin or ceramic to achieve optimal shade matching.3,6
MANUFACTURING AND PROCESSING TECHNIQUES
PEEK restorations and frameworks are fabricated through several pathways. Subtractive CAD/CAM milling of pre-polymerised, industrially manufactured PEEK blanks is currently the most widely used and clinically validated method, offering consistent material properties and minimal internal porosity compared with other techniques.2,3 Injection moulding and hot-pressing from granules or pellets represent alternative conventional fabrication routes, particularly for removable prosthodontic frameworks.3
Additive manufacturing (3D printing) of PEEK via fused filament fabrication, selective laser sintering, or related high-temperature extrusion processes is an emerging but technically demanding option, owing to PEEK's high melting temperature and the need for tightly controlled build-chamber conditions to avoid warping, voids, and inconsistent crystallinity. While promising for complex, patient-specific geometries such as maxillofacial prostheses and hybrid full-arch frameworks, additively manufactured PEEK is not yet as extensively validated clinically as milled PEEK.3
PEEK IN FIXED PROSTHODONTICS
In fixed prosthodontics, PEEK and its reinforced variants have been investigated as substructure material for single crowns, three- to four-unit fixed dental prostheses (FDPs), and post-and-core systems, typically veneered with composite resin to improve esthetics.6 The lower elastic modulus is proposed to act as a shock absorber beneath a veneering material, potentially reducing chipping and debonding compared with rigid ceramic frameworks in select situations, though this benefit must be weighed against PEEK's lower intrinsic flexural strength, which restricts its indication largely to short-span, low-to-moderate load-bearing situations rather than long-span posterior FDPs.5,6 PEEK has also been explored as an endodontic post material, capitalising on its bone/dentin-like modulus to reduce the risk of root fracture compared with rigid metal or fibre posts, although bonding to radicular dentin remains a technical challenge discussed further below.
PEEK IN REMOVABLE PROSTHODONTICS
PEEK has attracted considerable interest as a metal-free framework and clasp material for removable partial dentures (RPDs), offering a lighter, more esthetic, and hypoallergenic alternative to cast cobalt-chromium frameworks for patients with metal sensitivity or esthetic concerns in the clasp-visible zone.7,8 In-vitro fatigue and retention studies, along with recent systematic reviews, generally report that PEEK clasps demonstrate retentive forces and cyclic fatigue resistance that are comparable to, and in some testing conditions superior to, conventional Co-Cr clasps, while offering improved flexibility that may reduce abutment tooth stress during insertion and removal.10 However, PEEK RPD frameworks demonstrate lower rigidity than Co-Cr in major connector regions, which can necessitate design modifications (increased bulk or reinforcement) to maintain adequate resistance to flexure under function.7
For complete dentures, PEEK has been proposed as a denture base material, particularly for patients allergic to residual monomer in polymethyl methacrylate (PMMA) or those requiring a lighter prosthesis, and as a framework material for implant-retained overdenture bars, where its damping properties may help protect the underlying implants from peak occlusal loads.8,9 Clinical reports describing patient-reported outcomes are encouraging, but comparative long-term trials against conventional PMMA and metal-reinforced bases remain limited.
PEEK IN IMPLANT PROSTHODONTICS
Implant prosthodontics represents perhaps the most actively researched application of PEEK, spanning provisional and definitive abutments, implant-supported crown and FDP frameworks, and full-arch hybrid (all-on-four/all-on-six type) prostheses.4,5,11 Finite element analyses consistently demonstrate that PEEK abutments and PEEK-based restorations reduce peak stress concentration at the implant-abutment and implant-bone interfaces compared with rigid titanium or zirconia counterparts, supporting the theoretical rationale for reduced stress-shielding and, potentially, improved long-term marginal bone preservation.5
Systematic reviews of complete-arch implant-supported fixed prostheses fabricated from PEEK and the related polymer polyetherketoneketone (PEKK) report favourable short- to medium-term clinical performance, with technical complications predominantly veneer chipping or fracture rather than framework failure representing the most common adverse event. Early protocols for systematic reviews of PEEK hybrid prostheses in all-on-four rehabilitation likewise highlight promising implant survival and patient satisfaction, while noting that robust, long-term comparative data against titanium and zirconia hybrid frameworks are still awaited.11 PEEK has also been used as an interim material for immediately loaded implant provisional restorations, leveraging its machinability, adequate short-term strength, and radiolucency for uncomplicated postoperative radiographic assessment.6
SURFACE MODIFICATION AND BONDING CONSIDERATIONS
A well-recognised limitation of PEEK is its chemically inert, low-surface-energy, hydrophobic nature, which results in poor micromechanical and chemical bonding to resin cements and veneering composites unless the surface is appropriately conditioned.12 A range of surface treatments has been investigated, including airborne-particle abrasion with alumina, sulfuric acid etching (including piranha-solution etching combining sulfuric acid and hydrogen peroxide), plasma treatment, and laser irradiation.9,12 Across the available scoping and systematic reviews, sulfuric acid etching is the most consistently reported method for achieving the highest bond strength, by creating a porous, micro-retentive surface and introducing reactive functional groups, followed by alumina-particle air abrasion as a somewhat less aggressive but clinically more practical alternative.12 With regard to adhesive systems, resin cements or bonding agents containing methyl methacrylate (MMA) monomers, pentaerythritol triacrylate (PETIA), and dimethacrylates have demonstrated superior bonding performance to PEEK compared with conventional composite-based resin cements alone.12 Hydrofluoric acid etching, effective for glass-ceramics, has not shown comparable efficacy on PEEK surfaces owing to the absence of a glassy phase for the acid to selectively dissolve.12
Clinically, these findings translate into a practical protocol combining mechanical roughening (sandblasting or acid etching) with an MMA-based or specifically formulated PEEK-compatible primer/adhesive prior to resin cementation or composite veneering, since bond strengths achieved without any surface pretreatment are generally regarded as clinically insufficient for long-term retention.9,12
BIOCOMPATIBILITY AND BIOLOGICAL CONSIDERATIONS
PEEK is widely reported to be biologically inert, non-cytotoxic, and non-mutagenic, with a long track record of biocompatibility from its orthopaedic applications that has informed its adoption in dentistry.1,2 Its low surface energy and smooth, non-porous surface (in the unmodified state) are associated with reduced bacterial adhesion and plaque accumulation compared with some conventional prosthodontic materials, a property of particular relevance to peri-implant soft-tissue health and RPD-associated denture stomatitis.4,6 Unlike metallic alloys, PEEK does not release corrosion products or metal ions, making it a reasonable option for patients with documented metal hypersensitivity.4
A recognised biological limitation, however, is that unmodified PEEK is relatively bioinert and does not inherently promote osseointegration or direct bone bonding in the manner of roughened titanium surfaces, which has prompted research into bioactive surface coatings (e.g., titanium or hydroxyapatite coatings, plasma-sprayed layers) to enhance osteoconductivity when PEEK is used for implant fixtures themselves rather than superstructures.1,4 Long-term histological and clinical osseointegration data for PEEK dental implants remain considerably more limited than for titanium, and at present PEEK implant fixtures are not considered an evidence-based substitute for titanium implants; the stronger evidence base supports PEEK's use as an abutment, framework, or superstructure material rather than as the implant body itself.4
Summary of Current Clinical Evidence
Table 2 summarises the general direction of evidence across the principal prosthodontic applications of PEEK discussed above, drawing on the systematic reviews, clinical reports, and narrative reviews cited in this article. Readers are encouraged to consult the primary sources for detailed methodology, risk-of-bias assessment, and quantitative outcomes before drawing definitive clinical conclusions.
Table 2. Summary of current evidence for PEEK applications in prosthodontics
Application |
Evidence type |
Key finding (summarised) |
Ref. |
|
RPD frameworks / clasps |
Systematic reviews, in-vitro fatigue studies |
Comparable or superior retentive fatigue resistance to Co-Cr; more esthetic, lighter; long-term clinical data still limited |
10, 4 |
|
Implant-supported FDP frameworks (PEEK/PEKK) |
Systematic review of clinical studies |
Favourable short-to-medium-term survival; veneering chipping remains the main complication |
— |
|
Implant abutments / provisional restorations |
Narrative reviews, FEA studies |
Reduced peri-implant stress concentration due to bone-like modulus; adequate short-term performance as provisional/definitive abutments |
5, 6 |
|
All-on-four hybrid prostheses |
Systematic review protocols, clinical reports |
Encouraging patient satisfaction and reduced prosthesis weight; comparative long-term titanium data still awaited |
11 |
|
Removable complete denture bases / obturators |
Case reports, narrative reviews |
High patient comfort, allergy-friendly alternative to PMMA/metal; limited long-term outcome data |
8, 9 |
|
Maxillofacial / craniofacial prostheses |
Case series, patient-satisfaction surveys |
Good esthetic and functional outcomes reported for craniofacial defect rehabilitation |
— |
LIMITATIONS AND CHALLENGES
Despite encouraging short- to medium-term data, several limitations temper enthusiasm for widespread adoption of PEEK in routine prosthodontic practice. First, the majority of available evidence derives from in-vitro mechanical and bond-strength studies, finite element analyses, and small case series or retrospective clinical reports; prospective randomised controlled trials with extended follow-up (five years or more) remain scarce across nearly all applications discussed above.3,4 Second, esthetic limitations of unveneered PEEK necessitate composite or ceramic veneering in visible zones, introducing an additional interface prone to chipping and debonding, particularly given the bonding challenges described earlier.6,12 Third, the lower flexural strength of unfilled PEEK restricts its use in high-load, long-span posterior applications unless reinforced, and reinforcement strategies (carbon-fibre or ceramic filling) each carry their own trade-offs in esthetics or evidence maturity. Finally, standardised, material-specific clinical guidelines and long-term outcome registries comparable to those available for titanium implants or zirconia restorations have yet to be firmly established for PEEK, which complicates evidence-based case selection for the practising prosthodontist.3
FUTURE PERSPECTIVES
Several developments are likely to shape the future role of PEEK in prosthodontics. Continued refinement of ceramic- and fibre-reinforced PEEK composites may narrow the gap between esthetic performance and mechanical reliability, potentially expanding indications to longer-span fixed prostheses. Advances in additive manufacturing specific to high-temperature thermoplastics may improve the accessibility and geometric complexity achievable with PEEK, particularly for patient-specific maxillofacial and hybrid full-arch prostheses.3 Surface-functionalisation research including bioactive coatings, plasma-based surface activation, and novel primer chemistries is likely to further improve both bonding reliability for veneering and, where relevant, osteoconductivity for implant-related applications.1,9,12 Finally, the increasing number of ongoing and recently registered systematic review protocols in this area signals a maturing evidence base; well-conducted prospective clinical trials with standardised outcome reporting will be essential to move PEEK from a promising alternative material to a routinely evidence-supported option in prosthodontic practice.10,11
CONCLUSION
PEEK offers a biomechanically and biologically attractive profile a bone-like elastic modulus, chemical inertness, radiolucency, and reduced weight that supports its emerging role across fixed, removable, and implant prosthodontics. Current evidence, while encouraging, is predominantly short- to medium-term and weighted towards in-vitro and observational clinical data. For the prosthodontic postgraduate and clinician, PEEK should presently be regarded as a valuable, evidence-informed addition to the material armamentarium for selected indications particularly in patients with metal hypersensitivity, esthetic RPD clasp requirements, and situations where stress-shielding reduction is desirable rather than a wholesale replacement for well-established metallic and ceramic prosthodontic materials. Rigorous, adequately powered, long-term clinical trials remain the key requirement before PEEK's role can be more definitively established.
Conflict of Interest: Nil.
Source of Support: Nil.
REFERENCES
- Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials. 2007;28:4845-69.
- Panayotov IV, Orti V, Cuisinier F, Yachouh J. Polyetheretherketone (PEEK) for medical applications. J Mater Sci Mater Med. 2016;27:118.
- Schwitalla A, Müller WD. PEEK dental implants: a review of the literature. J Oral Implantol. 2013;39:743-9.
- Mishra S, Chowdhary R. PEEK materials as an alternative to titanium in dental implants: a systematic review. Clin Implant Dent Relat Res. 2019;21:208-22.
- Lee WT, Koak JY, Lim YJ, Kim SK, Kwon HB, Kim MJ. Stress shielding and fatigue limits of poly-ether-ether-ketone dental implants. J Biomed Mater Res B Appl Biomater. 2012;100:1044-52.
- Papathanasiou I, Kamposiora P, Papavasiliou G, Ferrari M. The use of PEEK in digital prosthodontics: a narrative review. BMC Oral Health. 2020;20:217.
- Zoidis P, Papathanasiou I, Polyzois G. The use of a modified poly-ether-ether-ketone (PEEK) as an alternative framework material for removable dental prostheses: a clinical report. J Prosthodont. 2016;25:580-4.
- Bathala L, Majeti V, Rachuri N, Singh N, Gedela S. The role of polyether ether ketone (PEEK) in dentistry - a review. J Med Life. 2019;12:5-9.
- Skirbutis G, Dzingut? A, Masili?nait? V, Šulcait? G, Žilinskas J. A review of PEEK polymer's properties and its use in prosthodontics. Stomatologija. 2017;19:19-23.
- Al-Banyahyati I, Sefrioui A, Berrada S, Zaoui F, Benyahia H. Retentive value and cyclic fatigue resistance of polyetheretherketone clasps as an alternative to cobalt-chrome clasps for removable partial dentures: a systematic review. Cureus. 2025;17:e98435.
- Boukhris H, Hajjami H, Ben Youssef S. Clinical outcomes of polyetheretherketone (PEEK) hybrid prosthesis in all-on-four rehabilitation: a systematic review protocol [version 1]. F1000Res. 2024;13:507.
- Machado PS, Rodrigues ACC, Chaves ET, Susin AH, Valandro LF, Pereira GKR, et al. Surface treatments and adhesives used to increase the bond strength between polyetheretherketone and resin-based dental materials: a scoping review. J Adhes Dent. 2022;24:337-56.
Disclaimer from Guident
The views and opinions expressed in articles published in Guident are solely those of the authors and do not necessarily reflect the views of the editors, the editorial board or the publisher, Ivory India. Guident does not endorse any product, technique or treatment described. Clinical content is intended for qualified dental and medical professionals and is not a substitute for individual clinical judgement. Guident, its editors and the publisher accept no liability for any consequence arising from the use of information published in the journal.
How to cite this article
Dr. Razousinuo Khezhie, Dr. Atoshi Pandit, Dr. Manesh Lahori. PEEK as an Alternative Biomaterial in Prosthodontics: Current Evidence and Future Perspectives. Guident. 2026 Aug;19(9). Available from: https://ivoryindia.in/articles/prosthodontics/peek-as-an-alternative-biomaterial-in-prosthodontics-current-evidence-and-future-perspectives.html