Comparative Evaluation of Flexural Strength and Hardness of 3D Printed and Conventional Heat Cure Denture Base Resins: An in Vitro Study
Department of Prosthodontics and Crown & Bridge, Government Dental College and Hospital Jamnagar, Gujarat, India
ABSTRACT
Context: This in vitro study compares the flexural strength and hardness of 3D printed and heat-cured denture base resins to determine the superior material for clinical use.
Aims: The aim of the study is to compare the flexural strength and hardness of 3D printed and heat-cured denture base resins to identify the superior material for clinical use.
Settings and Design: To measure and compare the flexural strength and hardness of 3D printed denture base resin (Prevest DenPro) and Lucitone heat-cured denture base resins.
Methods and Material: According to ISO 20795-1:2013 standards, 20 rectangular and 20 circular specimens were fabricated for each group: Group A (Lucitone denture base resin) and Group B (3D printed denture base resin). Mechanical testing was performed to assess flexural strength using a universal testing machine and hardness using the Vickers hardness test.
Statistical analysis used: SPSS 24.0 Software, and the student t test to analyse statistical significance.
Results: The mean flexural strength of Lucitone heat-cured denture base resin was 92.6 MPa, while for 3D printed resin, it was 121.05 MPa. The mean microhardness was 16.77 HV for Lucitone and 19.48 HV for 3D printed resin. Using the Student's t-test, the p-value was <0.0001, indicating the results were statistically significant.
Conclusions: 3D printed denture base resins exhibit superior flexural strength and hardness compared to conventional heat-cured resins, suggesting they may offer enhanced durability for clinical use.
Key-words: 3D printed denture base resin, Heat cured denture base resin, flexural strength, microhardness.
INTRODUCTION
Prosthodontic rehabilitation with complete dentures relies heavily on the mechanical and physical properties of the denture base material. Polymethyl methacrylate (PMMA) has remained the gold standard for conventional denture base fabrication for decades due to its acceptable aesthetics, ease of manipulation, and biocompatibility. However, limitations such as polymerization shrinkage, porosity, and suboptimal mechanical properties have prompted the search for alternative fabrication techniques and materials.1
The advent of additive manufacturing, particularly 3D printing, has introduced novel denture base resins that offer advantages including digital precision, reduced chairside time, and cost-effectiveness in long-term applications.2 These materials, often cured by light or UV polymerization, claim improved homogeneity and dimensional accuracy. Despite these benefits, concerns remain regarding their clinical longevity, particularly under masticatory stress and thermal changes. Therefore, properties like flexural strength and surface hardness become critical parameters to assess before 3D printed resins can be considered equivalent or superior to conventional PMMA.3
Additive manufacturing has opened a new era in prosthodontics, but the clinical success of 3D printed resins still depends largely on their mechanical behaviour.
This study aims to compare the flexural strength and hardness of conventional heat-cure PMMA denture base resins with that of 3D printed denture base materials in an in vitro setting, thereby evaluating their suitability for long-term clinical use.
MATERIALS AND METHODS
This in-vitro experimental study was designed to compare the flexural strength and microhardness of a 3D printed denture base resin (Prevest Denpro) with a conventional heat-polymerized denture base resin (Lucitone 199, Dentsply Sirona). The study was conducted according to ISO 20795-1:2013 standards for denture base resin testing.
Sample Preparation
A total of 40 samples were fabricated, divided equally into two groups:
Group A: Heat-cured PMMA resin (Lucitone 199)
Group B: 3D printed resin (Prevest Denpro)
Each group consisted of:
- 20 rectangular specimens (64 mm × 10 mm × 3.3 mm) for flexural strength testing
- 20 circular specimens (10 mm diameter × 2 mm thickness) for microhardness testing
Group A: Heat-Cure Denture Base Resin
- Rectangular and circular wax patterns were fabricated.
- Wax patterns were flasked, dewaxed, and packed with Lucitone 199 resin.
- Curing was carried out using conventional compression moulding technique.
- Post-polymerization, the samples were finished, polished, and stored in distilled water at 37°C for 24 hours before testing.
Group B: 3D Printed Denture Base Resin
- Samples were digitally designed using NX CAD software.
- STL files were generated and printed using the ASIGA Max 385 3D printer with a 50 µm layer thickness and 90° orientation.
- Post-processing included alcohol rinsing and UV light post-curing for 45–60 minutes as per the manufacturer’s recommendations.
- The final printed specimens were finished, polished, and stored similarly to Group A.
FLEXURAL STRENGTH TESTING
Flexural strength was evaluated using a 3-point bending test on a universal testing machine (Model STS-248, Star Testing Systems, India).
- Span length: 50 mm
- Crosshead speed: 5 mm/min
- Load applied until fracture
- Flexural strength (MPa) was calculated using the formula:
MICROHARDNESS TESTING
- Microhardness was measured using a Vickers hardness tester (Reichart, Austria).
- A load of 50 g was applied for each indentation.
- Two indentations were made per specimen, 1 mm apart, and the average value was taken as the hardness for that specimen.
- Hardness was expressed in Vickers Hardness Number (VHN).
RESULTS
A total of 40 samples were analyzed, with 20 specimens in each group. Group A included heat-polymerized denture base resin specimens (Lucitone 199), and Group B comprised 3D printed denture base resin specimens (Prevest Denpro). The mechanical properties flexural strength and Vickers microhardness were assessed and compared between the two groups.
FLEXURAL STRENGTH
The mean flexural strength of Group A (Lucitone 199) was 92.6 MPa, while Group B (Prevest Denpro) exhibited a significantly higher mean flexural strength of 121.05 MPa. The difference between the groups was statistically significant (p < 0.05), indicating superior resistance to bending forces in the 3D printed specimens.
VICKERS MICROHARDNESS
The mean microhardness value recorded for Group A was 16.77 VHN, whereas Group B demonstrated a higher mean value of 19.48 VHN. The difference in hardness between the two groups was statistically significant (p < 0.05), suggesting greater resistance to surface indentation in the 3D printed material.
STATISTICAL ANALYSIS
Data were analyzed using an independent t-test. Both mechanical properties flexural strength and microhardness showed statistically significant differences in favour of the 3D printed denture base resin group, with p-values < 0.05.
DISCUSSION
The present study was conducted to evaluate and compare the flexural strength and microhardness of 3D printed denture base resin (Prevest Denpro) and conventional heat-polymerized PMMA resin (Lucitone 199). The results demonstrated that the 3D printed denture base resin exhibited significantly higher flexural strength and hardness than the conventional heat-cure material.
Flexural strength is a critical property for denture bases as it reflects the material's ability to withstand occlusal forces and prevent midline fractures. In this study, the mean flexural strength of Prevest Denpro was 121.05 MPa, which was significantly greater than 92.6 MPa for Lucitone 199. This suggests that the 3D printed resin is more resilient to bending and masticatory stress, which is in agreement with findings from Gad et al. 4 who reported improved flexural properties in 3D printed resins subjected to optimized printing parameters and post-curing protocols.
The enhanced strength in 3D printed specimens may be attributed to factors such as reduced porosity, layer-by-layer polymerization, 90° build orientation, and 50 μm layer thickness, which together facilitate a high degree of monomer conversion and interlayer bonding. Moreover, post-curing under UV light contributes to improved polymer cross-linking, further reinforcing the mechanical structure of the resin.5
Regarding microhardness, the 3D printed resin also showed significantly higher Vickers hardness values (19.48 VHN) compared to the conventional resin (16.77 VHN). Hardness is a key determinant of surface wear resistance, which influences long-term durability and the ability of the material to maintain polish and resist deformation. A higher degree of polymer conversion in 3D printed resins has been linked to increased surface hardness, as confirmed by studies conducted by Al-Dulaijan et al.6
However, the results of this study are not in agreement with the findings of Prpic et al.7, who reported lower flexural strength in 3D printed denture base resins compared to conventional resins. Such discrepancies could be due to differences in material brand, printing orientation, post-curing protocols, or layer thickness. This highlights the sensitivity of mechanical properties to the specific 3D printing parameters used.
It is important to acknowledge the in vitro nature of the present study, which may not fully replicate the oral environment. Factors such as thermal cycling, salivary enzymes, and long-term fatigue loading were not simulated, which may influence material performance clinically. Moreover, while 3D printed resins exhibit promising mechanical behavior, their long-term biological properties, repairability, and cost-effectiveness require further investigation before routine clinical implementation.
REFERENCES
- Anusavice KJ, Shen C, Rawls HR. Phillips' Science of Dental Materials. 12th ed. St. Louis: Elsevier; 2013.
- Revilla-Leon M, Ozcan M. Additive manufacturing technologies used for processing polymers: current status and potential application in prosthetic dentistry. J Prosthodont Res. 2019;63:26–34.
- Al-Dulaijan Y, Alsulaimi S, Alotaibi F, Alawadh A, Vohra F, Abduljabbar T. Comparison of flexural strength and hardness of 3D-printed, CAD/CAM milled, and heat-polymerized denture base resins. J Prosthodont. 2021; 30:566–572.
- Gad MM, Abualsaud R, Rahoma A, Al-Thobity AM. Strength and Surface Properties of a 3D-Printed Denture Base Polymer. J Prosthodont. 2022; 31:412–418.
- Reymus M, Stawarczyk B, Kessler A, et al. 3D-Printed Material for Temporary Restorations: Impact of Print Layer Thickness and Post-Curing Method. Int J Comput Dent. 2019, 22:231–237.
- Al-Dulaijan YA, Al-Shehri AA, et al. Comparative Evaluation of Surface Roughness and Hardness of 3D Printed Resins. Polymers (MDPI). 2022; 15:6822.
- Prpic V, Schauperl Z, Catic A, Dulcic N. Comparison of Mechanical Properties of 3D-Printed, CAD/CAM, and Conventional Denture Base Materials. J Prosthodont. 2020; 29:524–528.
- Shim JS, Kim JE, Jeong SH, Choi YJ, Ryu JJ, Lee KW. Printing accuracy, mechanical properties, surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations. J Prosthet Dent. 2020; 124:468-475.
- Alp G, Murat S, Yilmaz B. Comparison of flexural strength of different CAD/CAM PMMA-based polymers. J Prosthodont. 2019; 28:491–495.
- Infante L, Yilmaz B, McGlumphy EA, Johnston WM. Fabricating complete dentures with CAD/CAM technology. J Prosthet Dent. 2014; 111:351-355.
- Al-Dwairi ZN, Al-Hiyasat AS, Baba NZ. A comparison of the surface properties of CAD/CAM and conventional PMMA. J Prosthodont. 2019; 28:452–457.
- Paulino MR, et al. Simplified versus traditional techniques for complete denture fabrication: a systematic review. J Prosthet Dent. 2015; 113:12–16.
- Unkovskiy A, et al. Objects build orientation, positioning, and curing influence dimensional accuracy and flexural properties of stereolithographically printed resin. Dent Mater. 2018; 34:324–333.
List of Abbreviations:
|
Abbreviation |
Definition |
|
CAD |
Computer aided designing |
|
CAM |
Computer aided manufacturing |
|
PMMA |
Polymethylmethacrylate |
Disclaimer from Guident
The views and opinions expressed in articles published in Guident are solely those of the authors and do not necessarily reflect the views of the editors, the editorial board or the publisher, Ivory India. Guident does not endorse any product, technique or treatment described. Clinical content is intended for qualified dental and medical professionals and is not a substitute for individual clinical judgement. Guident, its editors and the publisher accept no liability for any consequence arising from the use of information published in the journal.
How to cite this article
Dr. Sanjay Lagdive, Dr. Neha Dhruve, Dr. Sukram Rathava. Comparative Evaluation of Flexural Strength and Hardness of 3D Printed and Conventional Heat Cure Denture Base Resins: An in Vitro Study. Guident. 2026 Mar;19(4). Available from: https://ivoryindia.in/articles/research/comparative-evaluation-of-flexural-strength-and-hardness-of-3d-printed-and-conventional-heat-cure-denture.html