Factors Aff Ecting the Implant Abutment Interface
Department of Prosthodontics Yenepoya Dental College, Yenepoya University, Mangalore, India
Abstract
The long term success of an implant depends not only the on the stability of bone support for the implant, but a stable Implant abutment interface. Implantology has evolved from the conventional protocols suggested by Branemark to more challenging and predictable approach. The purpose of this article is to present a brief review of literature on the factors that influence the success and survival rates of dental implants and also to review the newer concepts in implantology. Google Scholar was used to search the English language literature. The key words used included Implant success, Implant stability, Implant –bone interface, crestal bone loss, implant design, implant surface and implant abutment interface most recent articles were reviewed and additionally references were obtained from a hand search of reviewed articles
INTRODUCTION
Rehabilitation of missing teeth using endosseous root form titanium dental implants has proven to be a predictable and highly successful treatment modality. The success of the treatment depends on many factors affecting the bone–implant, Implant–abutment and abutment–prosthesis interfaces.
A comprehensive review of the literature by Charles J. Goodacre (1), reported mechanical complications of implants such as the presence of fistula at the level of the abutment connection, , loss of abutment screws, loss of unit crowns; abutment screw fractures and mechanical failure due to implant fracture. These mechanical failures may be directly or indirectly attributed to a compromised implant abutment connection. In addition to these mechanical complications, improper connection may also cause certain biological complications like crestal bone loss and soft tissue inflammation due to micro movements or micro gaps at the implant abutment interface.
The purpose of this article is to present a brief review of literature on the factors that influence the Implant Abutment interface and thereby the success and survival rates of dental implants. Google Scholar was used to search the English language literature. The key words used included Implant success, implant stability, implant–bone interface, crestal bone loss, implant Abutment connection design, Micro motion and platform switching. The most recent articles were reviewed and additionally references were obtained from a hand search of reviewed articles.
FACTORS EFFECTING THE IMPLANT ABUTMENT INTERFACE
The structural integrity and biologic compatibility of the implant abutment interface and thereby that of the implant prosthesis depends on the following factors
- Implant abutment interface geometry/design
- Micromotion
- Screw mechanics
- Platform switching
IMPLANT ABUTMENT INTERFACE DESIGN
The implant abutment interface design plays a pivotal role in the stability and stress dissipation between the implant and abutment. External hex was the first implant abutment connection to be used but it had disadvantages like reduced prosthetic space due to the hexagonal projection from the implant surface. Loosening of abutment screws, particularly with the initial external-hex implant systems, has been a well-known technical problem, occurring mainly during the first 2 years after delivery of the crown.
Dr. Gerald A Niznick was the first one to suggest modification to the Implant abutment design in the form of internal hex. In internal hex connection the mating components are situated inside the implant body which was believed to help in better stress distribution and provide better and more prosthetic options. Almost all vitro studies2-5, with the exception of one,6 have demonstrated that internal connections have greater stability than external hex connection.
The next major advancement in terms of geometry of the connection was the introduction of tapered connections. Tapered connections are believed to give better marginal seal and reduce the micro movements between the implant and the abutment. A conical implant-abutment interface at the level of the marginal bone is also believed to improve the distribution of the load into the supporting bone.7 But titanium conical abutments appear to have poorer load fatigue performance compared with external-hexagon connections.8 In reduced-diameter conical connections, the neck of this implant is a potential zone for fracture when subjected to high bending forces.9
The joined surfaces may also incorporate a rotational resistance and indexing feature and/or lateral stabilizing geometry. This geometry is further described as octagonal, hexagonal, cone screw, cone hex, cylinder hex, spline, cam, cam tube, and pin/slot. Combination of morse taper with other features such as internal hex is also being tried in newer implant systems. Some screw less implant systems which rely entirely on the friction fit for their stability are also available – Eg: Bicon.
MICROGAP AND MICROMOTION
When an abutment is connected to an implant fixture, there is a small space created between them, this is known as microgap. Studies have reported that the mean size of micro gaps was between 2 and 6 µm. Three main factors have been identified as possible causes for the formation of microgaps: occlusal load during physiological function,10 manufacturing tolerance11 and micromotion between the implant–abutment connections. Microorganisms occupy this gap and set up a bacterial reservoir, resulting in an area of inflamed soft tissue facing the fixture abutment junction.
The scanning electron microscopic analysis was used by Binon12, 1995 and Jansen et al13, 1997 to measure the gap between the abutment and implant. Binon12, analyzed the accuracy of the implant/abutment interface on ScrewVent implants with raises of 20, 100 and 200x and encountered a marginal gap of 45μm.
Micro motion in the system leads to cracks in the material at the implant surface as well as at the abutment and screw joint interfaces which leads to loss of preload and subsequently screw loosening. The closing of implant abutment screws at manufacturers’ recommended closing torque could help to minimize microgaps and the potentially adverse effects of micro leakage.14
SCREW MECHANICS
Almost all implant abutment connections are retained and stabilized by screws. A screw is a mechanism that converts rotational motion to linear motion, and torque (rotational force) to a linear force. .Preload is the technical term for the tension caused by tightening the screw that holds the assembled parts together. As long as the external loads on a joint don't exceed the preload, the screw is not subjected to any motion and will not become loose.
Screw loosening is the most common cause for failure of implant abutment connections15. Both over torqueing and under torquing the screw can lead to screw loosening. Over-torquing might cause failure by damaging the threads and deforming the fastener, though this can happen over a very long time. Under-torquing can cause failures by allowing a joint to come loose, and it may also allow the joint to flex and thus fail under fatigue. The most common mode of failure is overloading.
The two most probable causes of abutment screw loosening are excessive bending on the screw joint and settling effect of the joint. Excessive bending forces can either be on static or dynamic loads. Static loads greater than yield strength of screw causes deformation of the screw which leads to loss of preload and opening of interface. Dynamic loads leads to dynamic fatigue of the material. Settling effects represents the Nonabrasive wear of contact areas. When the total settling effect is greater than the elastic elongation of the screw, opposing forces no longer exist between the surfaces to hold the screw abutment stack together.
PLATFORM SWITCHING
Platform switching is the use of smaller diameter abutments on wider diameter implants. It is used as a method of reducing crestal bone loss and maintaining the gingival papillae. The implant abutment junction is displaced horizontally inwards from the perimeter of the implant platform, and further away from the bone. This creates an angle, or step, between the abutment and the implant. The inflamed connective tissue does not extend laterally to the same extent as it does with a traditional matched implant-abutment junction. Conical internal connections have what is called an in– built platform switch. As the abutment emerges out of the implant, the interface lies within the platform of the implant fixture. This itself will displace the inflammatory substances away from the crestal bone.
CONCLUSION
According to this brief literature review, one can conclude that with a conical implant–abutment interface at the level of the marginal bone, in combination with retention elements at the implant neck, and with suitable values of implant wall thickness and modulus of elasticity, the amount of micromotion can be reduce to a greater extend thereby improving joint stability and success of the implant therapy. The tightening torque is an important factor to improve mechanical and biological properties of the interface between implant and abutment and the use of the torque recommended from the manufacturer may potentially reduce the adverse effects of microleakage. The concept of platform switching appears to limit crestal resorption and seems to preserve peri-implant bone levels.
REFERENCES
- Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYK. Clinical complications with implants and implant prostheses. J prosthet dent. 2003 Aug;90(2):121–32
- Merz BR,Hunenbart S, Belser UC.Mechanics of the implant- abutment connection:An 8-degree taper compared to a butt- joint connection.Int J Oral Maxillofac Implants 2000;15: 519–526.
- Balfour A, O’Brien GR.Comparative study of antirotational sin- gle tooth abutments. J Prosthet Dent 1995;73:36–43.
- Norton MR.An in-vitro evaluation of the strength of an inter- nal conical interface compared to a butt joint interface in implant design. Clin Oral Implant Res 1997;8:290–298.
- Weiss EI,Kozak D, Gross MD.Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194–199.
- Piermatti J,Yousef H, Luke A,Mahevich R,Weiner S. An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems.Implant Dent 2006;15: 427–432.
- Hansson S. Implant-Abutment Interface: Biomechanical Study of Flat Top versus Conical. Clin Impl Dent Rel Res 2000; 2: 33–41.
- Y. L. Seetoh. Load Fatigue Performance of Conical Implant-Abutment Connections J Oral Maxillofac Implants 2011;26:797–806
- Kivang Akga. Evaluation of the mechanical characteristics of the implant-abutment complex of a reduced-diameter morse-taper implant – finite element stress analysis Clin. Oral Impl Res. 2003; 14: 444-454
- .Steinebrunner L, Wolfart S, Bo¨bmann K, Kern M. In vitro evaluation of bacterial leakage along the implant–abutment interface of different implant systems. International Journal ofOral & Maxillofacial Implants 2005; 20:875–81.
- Shareef N, Levine D. Effect of manufacturing tolerances on the micromotion at the Morse taper interface in modular hip implants using the finite element technique. Biomaterials1996; 17:623–30.
- BINON, P. P. Evaluation of machining accuracy and consistencyof selected implants standard abutments, and laboratory analogs.Int J Prosthodont, v.8, n.2, p.162-78, 1995.
- JANSEN, V., K.; CONRADS, G.; RICHTER, E-J. Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants,v.12, n.4, p.527-40, 1997.
- Takuma, Yoshiyuki and Matsumura. Marginal Fit and Microgaps of Implant-abutment Interface with Internal Anti-rotation Configuration. Dental Materials Journal 27(1) :29ï¼34, 2008
- McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics.Dent Clin North Am. 1998 Jan;42(1):71-89.
Disclaimer from Guident
The views and opinions expressed in articles published in Guident are solely those of the authors and do not necessarily reflect the views of the editors, the editorial board or the publisher, Ivory India. Guident does not endorse any product, technique or treatment described. Clinical content is intended for qualified dental and medical professionals and is not a substitute for individual clinical judgement. Guident, its editors and the publisher accept no liability for any consequence arising from the use of information published in the journal.
How to cite this article
Dr. Hasan Sarfaraz, Dr. Faris Mohammed Shafi, Dr. Kamlakanth Shenoy. Factors Aff Ecting the Implant Abutment Interface. Guident. 2013 Mar;6(5). Available from: https://ivoryindia.in/articles/implantology/factors-aff-ecting-the-implant-abutment-interface.html